GD6900 Capacitance da Factor Factor Factor
GD6900 yana auna ƙarfin ƙarfin da kuma asarar dielectric (tgδ) na babban ƙarfin lantarki na kayan lantarki.Yana da tsarin da aka haɗa, ginanniyar gadar gwajin hasarar dielectric, samar da wutar lantarki mai sauƙin daidaitawa, haɓaka mai canzawa da SF6 daidaitaccen capacitor.
Babban tushen wutar lantarki yana samuwa ta hanyar inverter na kayan aiki, wanda ake amfani da shi don gwada abu na gwaji bayan haɓaka mai canzawa.Idan ƙara ƙoƙon rufewa da na'urar sarrafa zafin jiki, ana iya gwada asarar dielectric na mai mai rufewa.
●Nunin allo na LCD.
●Ma'aunin juriya na insulation: Tsarin gwajin juriya da aka gina a ciki, na iya aiwatar da juriya na insulation (IR), rabon sha (DAR), gwajin polarization index (PI).
●Guntun kalanda da babban ajiya a ciki.Ajiye sakamakon gwaji bisa ga tsari na lokaci, duba rikodin tarihin kuma buga sakamakon.
●Ana iya fitar da bayanan kayan aiki ta hanyar U disk.
●Yanayin gwaji da yawa, tare da yanayin babban ƙarfin lantarki, babban ƙarfin waje, daidaitaccen ciki, daidaitaccen waje, GST/UST, haɓakar kai.High ƙarfin lantarki (fiye da 10kV) dielectric asarar gwajin za a iya yi a cikin halin da ake ciki na waje misali waje high ƙarfin lantarki.
●Gwada cikakken CVT (Mai Canjin Wutar Lantarki) C1 da C2 asarar dielectric da ƙarfin aiki a lokaci guda.Hakanan gwada rabon canjin CVT da bambancin kusurwar ƙarfin lantarki.
●Ana iya auna asarar dielectric da ƙimar ƙarfin ƙarfin C0 a cikin babban ƙarshen CVT ta amfani da hanyar kariya ta baya.
●Siginar samfur mai saurin gudu.Inverter da da'irar samfur a ciki ana sarrafa lambobi.Ana daidaita wutar lantarki ta ci gaba.
●LCR atomatik ma'auni.Inductance, capacitance, juriya, kwana ana iya aunawa da nunawa.
●Kariya da yawa na jujjuyawar wutar lantarki na shigarwa, gajeriyar kewayawa fitarwa, wuce-wuta-ƙarfi, kan-na yau, zafin jiki, amintacce kuma abin dogaro.A halin yanzu, yana da aikin gwajin ƙasa, ba a ba da izinin haɓaka ƙarfin lantarki don kayan aikin da ba na ƙasa ba.
●Babu buƙatar wargaza gubar HV don auna asarar dielectric da ƙarfin CVT.
●Ana iya canza mitar zuwa 50Hz, 47.5Hz/52.5Hz, 45Hz/55Hz, 60Hz, 57.5Hz/62.5Hz, 55Hz/65Hz.
| Yanayin aiki | -15 ℃ ~ 40 ℃ | RH <80% | ||||||
| Ka'idar hana tsangwama | Juyawa akai-akai | |||||||
| Tushen wutan lantarki | AC 220V± 10%, 50/60Hz | Za a iya amfani da Generator. | ||||||
| Babban ƙarfin lantarki | AC 0.5KV ~ 10KV (12kV na zaɓi) | Kowane 0.1kV | ||||||
| Daidaito | 2% | |||||||
| Max.halin yanzu | 200mA | |||||||
| Iyawa | 2000VA | |||||||
| Ƙarfin tashin hankali | AC 0V~50V/15A | 50Hz, 60Hz mitar guda ɗaya 45HZ/55HZ 47.5HZ/52.5HZ 55HZ/65HZ 57.5HZ/62.5HZ Mitar biyu ta atomatik | ||||||
| Ƙaddamarwa | kashi: 0.001% | Saukewa: 0.001PF | ||||||
| Daidaito | ∆tgδ: ±(karanta *1.0%+0.040%) | |||||||
| ∆C x: ±(karanta *1.0%+1.00PF) | ||||||||
| Kewayon aunawa | tg ku | Ba tare da iyaka ba | ||||||
| C x | 15pF <Cx <300nF | |||||||
| 10KV | Cx <60 nF | |||||||
| 5KV | Cx <150 nF | |||||||
| 1KV | Cx <300 nF | |||||||
| gwajin CVT | Cx <300 nF | |||||||
| Ma'aunin LCR | L> 20H (2kV) | R>10KΩ(2kV) | ||||||
| daidaiton ma'aunin LCR | 0.1% | Ƙaddamarwa | 0.01 | |||||
| Farashin CVT | 10 ~ 10000 | |||||||
| Daidaitaccen rabon CVT | 0.1% | |||||||
| Ƙididdigar rabo na CVT | 0.01 | |||||||
| Juriya na rufi | DC high ƙarfin lantarki | 5-10kV | Daidaito | ± (2%* Karatu +10V) | ||||
| Daidaiton IR | 100kΩ-1000GΩ: <5% (gwajin gwajin ba kasa da 250V) 100GΩ-1000GΩ: 10% (gwajin gwajin ba kasa da 10000V) | |||||||
| Girma | Babban naúrar: 350(L)×270(W)×270(H) | |||||||
| Akwatin kayan haɗi: 350(L)×270(W)×160(H) | ||||||||
| Ƙarfin ƙwaƙwalwa | Ƙungiyoyi 200, ana tallafawa ma'ajin faifan filashin USB. | |||||||
| Nauyi | Babban sashi: 22.75Kg | |||||||
| Akwatin kayan haɗi: 5.25Kg | ||||||||
| Mai gwadawa | 1 saiti |
| Kayan kayan haɗi | guda 1 |
| HV jan gwajin gubar 8m | guda 1 |
| LV baki gwajin gubar 8m | guda 1 |
| Igiyar wutar lantarki 1.8m | guda 1 |
| CVT Haɗa gubar 5m | guda 1 |
| Gudun ƙasa 2m | guda 1 |
| Buga takarda | 2 rowa |
| 10 A fus | guda 3 |
| Jagorar mai amfani | 1 kwafi |
| Rahoton Gwajin Masana'antu | 1 kwafi |
| Na'urorin haɗi na zaɓi | |
| Kwayoyin Gwajin Mai | 1 pc |
| Mai kula da dumama | 1 pc |
Tare da tantanin gwajin mai da mai kula da dumama, mai amfani zai iya gwada tan delta na dielectric ruwa kamar mai canza wuta.
Mai kula da dumama
Jirgin mai