GDFJ-VI Transformer Narkar da Gas Analyzer
GDFJ-VI Mai Canjawa Narkar da Gas Analyzer shine chromatograph gas mai ɗaukuwa wanda ya dace da saurin bincike akan wurin.Yana haɗu da ganowa, bincike da ganewar asali a cikin ɗaya, da kuma micro detector, ƙaramin tushen iskar gas da ginanniyar kwamfutar allo mai taɓawa.Yana da haɗe-haɗe mafi girma, mafi kyawun kwanciyar hankali, mafi kyawun siffa da mafi dacewa aiki fiye da chromatography mai ɗaukar nauyi na gargajiya.Binciken kan yanar gizo na iya yin nazari akan lokaci da gano yanayin kuskuren canji, yin aikin wutar lantarki mafi sauƙi kuma mafi dacewa.Kayan aikin yana ɗaukar tsarin gano mai ganowa uku na gargajiya, kuma bayanan nazari iri ɗaya ne da na chromatograph na al'adar dakin gwaje-gwaje.Kayan aikin yana ɗaukar ingantaccen ra'ayi na ƙirar ƙirar ƙira.Yana da babban haɗin kai, ƙananan girman da nauyin nauyi, wanda yake da sauƙin ɗauka zuwa shafin don bincike.Yana ba da cikakkun bayanai masu dacewa da daidaitattun bayanai don sa ido da gwajin taswira, kuma yana adana yawancin ma'aikata da kayan aiki.
Cikakken bincike na abubuwan da ke cikin sassan gas guda bakwai H2, CO, CO2, CH4, C2H4, C2H6, C2H2(hydrogen, carbon monoxide, carbon dioxide, methane, ethylene, ethane, acetylene) narkar da a cikin insulating mai za a iya kammala a daya allura.Matsakaicin adadin gano acetylene shine 0.1ppm, kuma an samu nasarar amfani da shi ga masana'antu da sassa da yawa kamar kamfanonin samar da wutar lantarki, masana'antar wutar lantarki, masana'antar transfoma, manyan masana'antar narkewa, da sassan samar da wutar lantarki na layin dogo.
●Gina-in touch allo fasahar anti-control don gane gani gani na kayan aiki iska matsa lamba.
Ana nuna duk matsi na iska ta lambobi ta hanyar firikwensin matsa lamba.Matsin iskar gas mai ɗaukar nauyi, iskar hydrogen da iska za a iya karanta kai tsaye ta wurin aiki, kuma ana iya ɗaukar yanayin bincike a ainihin lokacin.
●Yin amfani da fasahar anti-control allon taɓawa don cimma saitunan zafin kayan aiki, nuni da sauran ayyuka.
Ana iya nuna duk sigogin zafin jiki a lambobi.Za'a iya karantawa da sarrafa zafin jiki na akwatin shafi, mai ganowa da mai gyarawa kai tsaye ta wurin aiki.Aikin yana da sauƙi kuma a bayyane.
●Danna allo don kammala dumama, sanyaya, kunnawa da saitunan yanzu.
Na'urar sarrafa hanyar iskar da aka gina ta atomatik tana cire wani ɓangare na iska ta atomatik lokacin aiki akan allon taɓawa, ta yadda rabon hydrogen zuwa iska ya kai ga ma'aunin da ya dace da kunnawa, kuma ya dawo da rabo bayan an gama kunnawa ta atomatik;Hakanan, ta taga saitin saitin gada a cikin software na aiki zai iya kammala saitin gadar thermal.
●Fasahar saitin sifili na dijital FID don haɓaka ƙarfin hana tsangwama na kayan aiki.
Mai gano FID yana amfani da fasahar saitin sifili na lantarki da madaidaicin da'irar dijital don rama ainihin halin yanzu na FID, ta yadda za a gane ka'idojin lantarki na matakin siginar fitarwa na FID.Domin ana iya shigar da ƙa'idar lantarki kai tsaye akan tashar sigina na allon kewayawa don haɓaka ƙarfin hana surutu na lantarki, babu siginar tsangwama lokacin da aka haɗa na'urar potentiometer na inji.Saboda rage sigogi na parasitic, yana da ƙarfin hana tsangwama.
●Amfani da fasahar hana sarrafa kwamfuta don cimma nasarar sarrafa oscillator.
Ta hanyar layin siginar haɗawa, yana iya ba da ƙarfi ga ƙaramin oscillator, kuma ta wurin aiki, yana iya aiwatar da dukkan tsarin oscillator, kamar dumama, fara oscillation, tsayawa, buzzing da sauri a ƙarshen, da dai sauransu ystem a ciki. Turanci da Sinanci, waɗanda za a iya canzawa kyauta.
●Ginin famfo iska mai shiru, ba tare da tushen iska na waje ba.
Tare da ƙaramin famfon iska da aka shigo da shi a cikin babban injin, zai iya samar da tushen iska don kayan aiki ba tare da katsewa ba.Saboda ƙirar ƙaƙƙarfan ƙira, famfo na iska na iya yin aiki cikin shiru, kuma mai amfani ba zai ji kasancewarsa ba.
●Aikin kariyar karyewar iskar iskar gas na thermal conductive waya tungsten.
Tare da aikin kariyar yankewar zafin zafi mai hankali, lokacin da iskar gas mai ɗaukar nauyi ya yi ƙasa da madaidaicin aminci, tsarin zai yanke ta atomatik na zafin zafin na yanzu don kare mai gano yanayin zafin zafi daga lalacewa.
| Bangaren | Min.ganewa taro |
| H2 | 2 |
| CO | 5 |
| CO2 | 10 |
| CH4 | 0.1 |
| C2H4 | 0.1 |
| C2H6 | 0.1 |
| Farashin C2H2 | 0.1 |
FID (mai gano ionization na harshen wuta)
●Iyakar ganowa: Mt≤3×10-12g/s (Hexadecane/isooctane)
●Tushen amo: ≤5×10-14A
●Matsala ta asali: ≤1×10-13A/30min
●Kewayon layi: ≥ 106
TCD (Mai gano zafin zafin jiki)
●Hankali: S≥3500mV•ml/mg (na al'ada);5000mV•ml/mg (m sosai)
●Tushen amo: ≤10μV
●Matsakaicin tushe: ≤30μV/30min
●Kewayon layi: ≥104
Fihirisar sarrafa zafin jiki
| Bangaren kula da yanayin zafi | Rage | Daidaito |
| Akwatin ginshiƙi | 4 ~ 450 ℃sama da zafin jiki | ± 0.1 ℃ |
| Injector shafi | 4 ~ 450 ℃sama da zafin jiki | ± 0.1 ℃ |
| Mai gano FID | 4 ~ 450 ℃sama da zafin jiki | ± 0.1 ℃ |
| TCDMai ganowa | 4 ~ 450 ℃sama da zafin jiki | ± 0.1 ℃ |
| Nickel mai kara kuzari garambawuler | 4 ~ 450 ℃sama da zafin jiki | ± 0.1 ℃ |
Wasu
Girma: 420*280*300mm
Nauyi: ˂15kg
Wutar lantarki: AC220V± 10%, 50Hz
| A'a. | Suna | Ƙayyadaddun bayanai | Qty |
| 1 | DGA mai ɗaukar nauyi | GDFJ-VI | 1saita |
| 2 | Chromatography wurin aiki | Insulating mai bugu na musamman | 1saita |
| 3 | Karamin Cikakken Oscillator Na atomatik | 1saita | |
| 4 | High Tsarkake Nitrogen | 2L, 99.999%, Aluminum gami kwalban | 2kwalabe |
| 5 | High Purity Hydrogen | 2L, 99.999%, Aluminum gami kwalban | 1kwalban |
| 6 | Daidaitaccen Gas | 2L, 7 Abubuwan, Aluminum gami kwalban | 1kwalban |