GDYB-D24 Tsarin Gwajin Mita Makamashi Guda Daya
●Zane mai ma'ana da kyakkyawan bayyanar.Firam ɗin tebur da ke rataye na'urar da firam ɗin jikin tebur duk an yi su ne da kayan gami na musamman na aluminum, kuma saman tebur ɗin an yi shi ne da kayan fiber mai hana wuta da zafi.Dukan na'urar tana da launuka masu laushi, daidaitattun daidaituwa, da ingantaccen tsari.Kyawawan sana'a, kyakkyawan bayyanar.
●Aikin madannai na sadaukarwa da aikin PC suna tare, mai sauƙin amfani.Wannan na’ura tana dauke da maballin madannai, maballin yana zuwa da nunin faifan kristal mai ruwa, kuma yanayin aikin na’urar a bayyane yake.Aikin yana da sauƙi kuma kyauta.Software na sarrafa PC yana da matakai masu haske, kyakkyawar mu'amala, da sassauƙa da amfani mai dacewa.
●Babban digiri na atomatik.Lokacin tabbatar da mita makamashin lantarki, kawai buƙatar shiga cikin agogon da hannu, daidaita samfurin photoelectric, da tsara sigogin mita daidai, duk abubuwan tabbatarwa (ciki har da preheating, gwajin rarrafe, gwajin farawa, ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙayyadaddun ƙima, da kuma ƙayyadaddun kuskure na asali) ) PC yana sarrafa ta atomatik kuma an kammala shi a lokaci ɗaya.Hanyar daidaitawa na iya zama daidai da ƙa'idodin ƙasa ko kuma mai amfani na iya keɓance shi.
●Mai ɗaukar hoto mai ɗaukar hoto zai iya gane daidaitawa mai girma uku, kuma yana iya yin samfurin siginar juyawa na mitar makamashin lantarki mai jujjuya.Hakanan zai iya karɓar siginar stroboscopic na mitar makamashin lantarki.
●Tsarin sarrafa kurakurai na iya karɓar siginar samfurin samfurin photoelectric, bugun fitarwa na mitar lantarki da aka ɗora da siginar bugun jini mai ƙarfi na daidaitaccen ƙarfin lantarki.
●Mita na makamashi iri ɗaya (ƙarfin lantarki iri ɗaya da na yau da kullun ana iya ɗaukar su a lokaci guda (iyakance ga lambobi biyu).
●Yana iya gano alamar baƙar fata ta atomatik na mitar makamashin lantarki mai juyawa, wanda ya dace don yin ruwa (farawa) gwaji.
●2 zuwa 21 masu jituwa za a iya sanya su a kan madaidaicin igiyar wutar lantarki, kuma ana iya saita girman da lokaci na masu jituwa kamar yadda ake buƙata.Hakanan yana iya yin bincike mai jituwa akan fitarwa kuma ya zana sigar igiyar ruwa.
●Na'urar tana da babban kwanciyar hankali, ingantaccen aminci da babban ikon fitarwa.Wutar lantarki (na yanzu) madauki na iya ɗaukar nauyin kowane yanayi.
●Cikakken gano kuskure, wuri, kariya da ayyukan ƙararrawa don guje wa lalacewar na'urar saboda kurakuran wayar da mutum ya yi ko aiki.
●Software na sarrafa PC kuma yana da ayyuka kamar cikakken bugu na rahoto, ƙididdiga na lissafi da kiyaye kadara.
Daidaiton tsarin: 0.05%, 0.1%.
●Wutar lantarki mai fitarwa: 220V, daidaitacce kewayon: (0% ~ 120%) x kewayon aunawa, daidaitacce fineness: fiye da 0.01%.
●Fitowa mataki na yanzu: 0.1A, 0.25A, 0.5A, 1A, 2.5A, 5A, 10A, 25A, 50A, 100A.daidaitacce kewayon: (0% ~ 120%) x ma'auni kewayon, daidaitacce fineness: fiye da 0.01%.
●Ƙarfin fitarwa: madauki irin ƙarfin lantarki: 300VA, madauki na yanzu: 500VA.
●Ƙarfafa ƙarfin fitarwa, halin yanzu da ƙarfi:
≤0.02%/120S PF=1 (Class 0.05)
≤0.05%/120S PF=1 (Class 0.1)
●Karɓar igiyoyin wutar lantarki da na yanzu: ≤0.5%
●Kewayon mitar: 45Hz-65Hz daidaita daidaito: 0.01 Hz
●Matsakaicin canjin lokaci: -180.0°~180.0° daidaita daidaito: 0.01°
●Matsayin kayan aiki
●Daidaitaccen Voltmeter: matakin ± 0.2%
●Ammeter daidaito matakin ± 0.5%
●Daidaitaccen mita mataki: ± 0.5°
●Daidaiton mitar mitar: ± 0.1Hz
●Matsakaicin mita don gwaji: guda 12
●Wutar lantarki: 220V± 10%, 50Hz (ko 120V± 10%, 60Hz)
●Matsakaicin amfani da wutar lantarki: 800VA
●Yanayin aiki: 20± 5℃ dangi zafi: ≤85%
●Girma da nauyi: 2000mm*720mm*1550mm 280Kg