GDYB-S3 Tsarin Gwajin Mitar Makamashi Na Mataki Uku (Mataki 3)
Daidaiton tsarin: 0.1%;
Sigar tushe mataki uku:
a.Wutar lantarki na fitarwa:
3*57.7VY 3*100V ∆ ;
3*220V Y 3*380V ∆;
Daidaitacce kewayon 0 ~ 120% ci gaba da daidaitawa
Lalacewar daidaitawa: mafi kyau fiye da 0.01%
b.Fitowar halin yanzu: 25mA, 0.1A, 1A, 5A, 25A,100A;
Daidaitacce kewayon 0 ~ 120% ci gaba da daidaitawa
Lalacewar daidaitawa: mafi kyau fiye da 0.01%
Mafi ƙarancin fitarwa na yanzu: 1mA (daidai: <5%)
c.Ƙarfin fitarwa: (tsarin wutar lantarki) 3 * 50VA
(da'ira na yanzu) 3 * 150VA
d.Wutar lantarki mai fitarwa, halin yanzu, kwanciyar hankali: ≤0.02%/120S
e.Matsakaicin ƙimar ƙarfin fitarwa da yanayin yanayin halin yanzu: ≤0.5%
f.Alamar kusurwoyi uku: tsakanin 120°±0.3°
g.Kewayon mitar: 45Hz ~ 65Hz, mataki: 0.01Hz
h.Matsayin kusurwar lokaci: -180°~ 180°, mataki: 0.01°
i.Yana nuna daidaiton mita: Voltmeter, ammeter: ± 0.01%
Mitar mataki: ± 0.01°
Mitar mitar: ± 0.01Hz
Mafi ƙarancin fitarwa na yanzu: 1mA;daidaito: ≤5%;
Ƙarfin lodin fitarwa (mita 20)
Ƙarfin wutar lantarki: 800VA;
fitarwa na yanzu: 2500VA;
Voltage & daidaitawa na yanzu
Daidaitaccen kewayon: (0% ~ 120%) x kewayon aunawa;
Daidaitacce tarar: 10%, 1%, 0.1%, 0.01%;
Daidaita mataki:
Daidaitaccen kewayon: 0 ° ~ 360 °;
Daidaitacce fineness: 10°, 1°, 0.1°, 0.01°;
Daidaita mita:
Daidaitaccen kewayon: 45Hz ~ 65Hz;
Daidaitacce tarar: 1Hz, 0.1Hz, 0.01Hz;
Ƙarfin ƙarfin fitarwa: ≤0.05% / 120S;
Fitar wutar lantarki & murdiya na yanzu:
Wutar lantarki: ≤0.5%;
Yanzu: ≤0.5%;
Saituna masu jituwa:
Lambar masu jituwa: 2 ~ 21 sau;
Abubuwan jituwa: na yanzu 2 ~ 21 sau ≤40% (dangane da raƙuman ruwa na asali);
ƙarfin lantarki 2 ~ 21 sau ≤40% (dangane da mahimmancin kalaman);
Harmonic lokaci: 0 ° ~ 359.99 °;
Kayan aikin sa ido:
Daidaitaccen nuni na mita: Ƙarfin wutar lantarki: 0.01%;
Yanzu: 0.01%;
Ƙarfin ƙarfi: 0.01%;
Mataki: 0.01;
Yawan: 0.01;
Keɓaɓɓen ma'aunin wutar lantarki na yanzu 1:1:
Daidaito: 0.01%;
Ƙarfin fitarwa: 60VA;
Ma'auni na yanzu: 1mA ~ 120A (1: 1);
Babban fihirisar ayyuka
a) Daidaito: 0.01% (COS∮=0.5C…1…0.5L);
b) Ƙididdigar mita: 50Hz/60Hz± 5%;
c) Babban halin yanzu: 3 × 0.001-120A;
d) Na biyu na yanzu: 3×0.001-120A;
e) Nauyin fitarwa (kowane lokaci):
0.5W (kewayon fitarwa: 0.001-1A);
0.7V ya raba ta ƙimar halin yanzu (kewayon fitarwa: 1-120A);
f) Wutar lantarki mai fitarwa: Max.0.5V (fitarwa na yanzu Max. 120A);
Girman mai canzawa na yanzu: 265 x 200 x 240 mm (L x W x H);
Nauyin: 12kg;
Girma:
Tushen sarrafawa: 610 x 800 x 1920mm (L x W x H);
Girman mita: 2350 x 750 x 1850mm (L x W x H);