GDYB-S20 Tsarin Gwajin Mita Makamashi Mataki na Uku
Jerin GDYB Uku Daidaita Mahimmanci Multi-Ayyukan Tsarin Gwajin Makamashin Makamashin Wutar Lantarki yana da ingantaccen ingantaccen tsarin lantarki na ƙarni na uku wanda zai iya tsara tsarin gwajin aikin mita uku na lantarki bisa ga sabbin ƙa'idodi da ƙa'idodi na ƙasa.Yana ɗaukar madaidaicin 0.01% babban madaidaicin 1: 1 fasahar taswira na yanzu, don cimma gwajin ƙarfin lantarki na ciki da aka haɗa shunt mita uku.Ya dace don gwada mitar makamashin lantarki na kashi uku na 0.2% da ƙasa (ya haɗa da mitar wutar lantarki ta sinusoidal).Yana da kyakkyawan bayyanar, cikakkun ayyuka da kyakkyawan aiki.
GB/T 11150-2001 Kayan gwaji don mita makamashin lantarki (neq IEC 60736)
JJG 307-2006 Mitar Electromechanical don Auna Madadin-madaidaicin Makamashin Lantarki na yanzu
JJG 596-1999 Mitar Makamashin Wutar Lantarki Tare da Kayan Lantarki
JJG 597-2005 Kayan Tabbatarwa don Mitar Makamashi na AC
DL/T 614-1997 Multifunction mita watthur
DL/T 614-2007 Multifunction lantarki metering kayan aiki (AC)
Yanayin muhalli
Yanayin zafin aiki mai aiki: 23 ℃ ± 2 ℃.
Yanayin zafi mai aiki: 45% ~ 75%.
Tushen wutan lantarki
Single lokaci: AC220V± 33V.Mitar: 50± 2.5Hz
Nau'in Lamba: GDYB-S20 Tsarin Gwajin Mitar Makamashi Na Wuta Uku.
Daidaiton tsarin: 0.05%.0.1%.
Madaidaicin mita daidai: 0.05% Mita mai aiki da yawa mai aiki da yawa.
Voltage & Kewayon aunawa na yanzu:
Matakin fitarwa: 57.7V/100V/220V/380V.
Fitowar mataki na yanzu: 0.005A, 0.025A, 0.1A, 0.25A, 0.5A, 1A, 2.5A, 5A, 10A, 25A, 50A, 100A.
Mafi ƙarancin fitarwa na yanzu: 1mA.daidaito: ≤5%.
Ƙarfin lodin fitarwa (mita 20)
Ƙarfin wutar lantarki: 800VA.
fitarwa na yanzu: 2500VA.
Voltage & daidaitawa na yanzu
Daidaitaccen kewayon: (0% ~ 120%) x kewayon aunawa.
Daidaitacce tarar: 10%, 1%, 0.1%, 0.01%.
Daidaita mataki:
Daidaitaccen kewayon: 0° ~ 360°.
Daidaitacce fineness: 10°, 1°, 0.1°, 0.01°.
Daidaita mita:
Daidaitaccen kewayon: 45Hz ~ 65Hz.
Daidaitacce tarar: 1Hz, 0.1Hz, 0.01Hz.
Ƙarfin ƙarfin fitarwa: ≤0.05%/120S.
Fitar wutar lantarki & murdiya na yanzu:
Wutar lantarki: ≤0.5%.
Yanzu: ≤0.5%.
Saituna masu jituwa:
Lambar jituwa: 2 ~ 21 sau.
Abubuwan jituwa: sau 2 ~ 21 na yanzu ≤40% (dangane da raƙuman ruwa na asali).
ƙarfin lantarki 2 ~ 21 sau ≤40% (dangane da mahimmancin kalaman).
Harmonic lokaci: 0°~359.99°.
Kayan aikin sa ido:
Daidaitaccen nuni na mita: Wutar lantarki: 0.01%.
Yanzu: 0.01%.
Ikon: 0.01%.
Shafin: 0.01.
Yawan: 0.01.
Keɓaɓɓen ma'aunin wutar lantarki na yanzu 1:1:
Daidaito: 0.01%.
Ikon fitarwa: 60VA.
Ma'auni na yanzu: 1mA ~ 120A (1: 1).
Babban fihirisar ayyuka
a) Daidaici: 0.01% (COS∮=0.5C…1…0.5L).
b) Ƙididdigar mitar: 50Hz/60Hz±5%.
c) Babban halin yanzu: 3×0.001-120A.
d) Na biyu na yanzu: 3×0.001-120A.
e) Nauyin fitarwa (kowane lokaci):
0.5W (kewayon fitarwa: 0.001-1A).
0.7V an raba ta ƙimar halin yanzu (kewayon fitarwa: 1-120A).
f) Wutar lantarki mai fitarwa: Max.0.5V (fitarwa Max. 120A).
Girman mai canzawa na yanzu: 265 x 200 x 240 mm (L x W x H).
Nauyi: 12kg.
Girma:
Tushen sarrafawa: 610 x 800 x 1920mm (L x W x H).
Girman mita: 2350 x 750 x 1850mm (L x W x H).