HV HIPOT GDC-9560B Tsarin wutar lantarki Insulation Oil Gas Chromatography Analyzer
Shigarwa da cire ginshiƙin chromatographic:
1. Dole ne a aiwatar da shigarwa da kuma cire ginshiƙi na chromatographic a dakin da zafin jiki.
2. Cikakkun hasumiyai suna da hatimin ferrule da hatimin gasket.Akwai nau'ikan ferrules guda uku: ferrules na ƙarfe, ferrules na filastik da graphite ferrules, waɗanda ba su da sauƙi a ɗaure su yayin sakawa.Makullin nau'in Gasket na buƙatar sabbin gaskets a duk lokacin da aka shigar da shafi.
3. Ko ƙarshen biyu na ginshiƙi na chromatographic an haɗa su da ulun gilashi.Hana ulun gilashi da tattarawa daga hura su cikin injin gano iska ta hanyar iskar gas.
4. Tsawon ginshiƙin ginshiƙai da shigarwa ya dogara da littafin koyarwa na kayan aiki.Tsarin ɗakuna daban-daban na evaporation na chromatographic ya bambanta, kuma tsayin shigarwa shima ya bambanta.Ya kamata a lura cewa idan an yi amfani da ginshiƙin capillary tare da kwarara maras tsagewa, tsaka-tsakin tsaka-tsakin da ke tsakanin ɗakin vaporization da ginshiƙan da aka cika bai kamata ya kasance yana da yawan bincike da aka haɗa da ɗakin da aka haɗe ba, kuma ginshiƙi na capillary ya kamata ya zama dan kadan fiye da hula.
Tasirin rabon hydrogen zuwa iska akan mai gano FID:
A cikin chromatograph gas, rabon hydrogen zuwa iska ya kamata ya zama 1:10.Idan rabon ya yi yawa, hankalin mai gano harshen wuta na hydrogen zai ragu sosai.Bincika adadin hydrogen da kwararar iska.Lokacin da iskar hydrogen da iskar ba ta isa ba, zai yi “bang” ya kunna wuta, sannan ya kashe wutar, yawanci idan kun kunna murhun wutar lantarki, sai a kashe shi, sannan a sake kunna shi, sannan a kashe wutar. hydrogen bai isa ba.
Yadda za a hana allurar allurar lankwasawa:
Yawancin masu farawa a cikin chromatography sukan lanƙwasa allura da kara saboda:
1. Tashar allura na chromatograph iskar gas tana da ƙarfi sosai, kuma ana murƙushe ta sosai a zafin ɗaki.Lokacin da yawan zafin jiki na ɗakin vaporization ya ci gaba da tashi, gaket ɗin silicone zai faɗaɗa kuma ya ƙara ƙarfi.A wannan lokacin, yana da wuya a shiga ta sirinji.
2. An makale allurar a cikin ɓangaren ƙarfe na tashar allurar idan wurin ba shi da sauƙi a samu.
3. An lanƙwasa sandar sirinji saboda ƙarfin yana da ƙarfi yayin allurar samfurin.Akwai rakiyar samfuri don shigo da chromatographic.Ba shi yiwuwa a lanƙwasa sandar sirinji na gargajiya lokacin amfani da taragon samfurin don allurar samfurori.
4. Tun da bangon ciki na sirinji ya gurɓace, tura igiyar allura don tanƙwara yayin allura.
5. Gas chromatograph dole ne ya kasance karko yayin allura.Idan kuna gaggawa, zaku lanƙwasa sirinji.Idan kun saba da wannan aikin, zai ƙare da sauri.
Lokacin aikawa: Juni-07-2022