Lokacin da ma'aunin juriya na madauki (wanda ake kira ma'aunin juriya) wanda aka tsara bisa ga ƙa'idar ƙira ta gargajiya da aka gwada a cikin filin, an gano cewa akwai matsala gama gari.Lokacin da da'irar haɗin wutar lantarki na mai gwadawa ba ta da kyau ko kuma buɗe da'ira, mai gwadawa kuma zai nuna ƙima, kuma waɗannan yanayi zasu faru:
HV HIPOT GDHL-200A/100A Gwajin Juriya ta atomatik
(1) Wutar lantarki a buɗe take, kuma babu wani tsangwama mai ƙarfi na filin lantarki a wurin gwajin.A wannan yanayin, tun da bambancin yanayin shigar da wutar lantarki ta amplifier shine ainihin 0, ƙimar gwajin da kayan aiki ke nunawa yana kusa da 0. Idan mai gwadawa yana da isasshen ƙwarewar gwajin filin, zai iya yin hukunci cewa layin gwaji mara kyau na ƙarfin kayan aiki kewaye ba al'ada ba ne kuma komawa zuwa ƙarfin kayan aiki.Bayan an cire hanyar gwaji mara kyau, ana iya samun sakamakon gwajin daidai na ƙarshe.
(2) Alamar wutar lantarki ba ta da kyau.A mafi yawan lokuta, tashoshi na na'ura mai ba da hanya tsakanin hanyoyin sadarwa za su samar da fim din oxide ko fim din mai a saman farfajiyar tashar bayan aiki na dogon lokaci.Lokacin da maƙallan gwajin wutar lantarki don masu jujjuyawar da'irar ke manne zuwa irin wannan tashoshi na tasha, mummunan hulɗa yana yiwuwa.Lokacin da shirin gwajin wutar lantarki da kansa shima ya samar da takamaiman juriyar lamba, ƙimar juriyar lamba ɗaya ce da ƙimar juriyar ciki na kewayen samfurin ƙarfin lantarki.zai yi tasiri mai tsanani akan sakamakon gwajin.
(3) Madaidaicin wutar lantarki yana buɗewa ko kuma lambar sadarwa ba ta da kyau (lokacin da kewayawa ya buɗe, za a iya ganin cewa juriyar lamba r1 ba ta da iyaka), kuma akwai tsangwama mai ƙarfi na lantarki a wurin gwajin, kamar motar bas mai rai. .A wannan lokacin, bas ɗin mai rai yana wucewa ta capacitor tare da iska a matsayin matsakaici, yana tsoma baki tare da layin gwajin wutar lantarki guda biyu na mai gwadawa.Saboda tsangwama, layukan samun wutar lantarki na ma'aunin juriya na madauki biyu ne.Yanayin bambance-bambancen ƙarfin lantarki yana bayyana a ƙarshe.
Lokacin aikawa: Mayu-23-2022